When using the TDR100 to measure slope stability or rock deformation, what are the input values for PC-TDR?

Although slope stability/rock mass deformation users will need to adjust the following PC-TDR default settings, these settings may provide a good starting point:

  • Cable VP = 1.
  • Waveform Average = 4.
  • Waveform Points = 251.
  • Waveform Start = 0.
  • Waveform Length = 50.00.

Note that the Probe Length and Probe Offset will not be used for a slope stability/rock mass deformation application.