Can multiple CS547A probes be wired to one data logger if there is one A547 interface for each probe?

Two CS547A sensors can be wired directly to one CR1000. One A547 interface is needed for each sensor, unless a multiplexer is used. A multiplexer switches all sensors through one A547 interface. 

The CS547A instruction manual offers this warning: "Wire the conductivity excitation and temperature excitation lines in different excitation ports or measurement errors will occur. If multiple CS547A/A547s are wired to one data logger, each conductivity excitation must use a separate excitation port. However, you can wire multiple temperature excitation lines into the same excitation port."